Reflections from the panel
I recently joined a panel discussion at the PDA Aseptic Manufacturing Excellence Conference 2026; which was held in Melbourne, Australia on 14th October 2026, on “The Great Isolator Debate” which certainly lived up to its name. Bringing together perspectives from manufacturing, equipment suppliers, and regulation made for a lively discussion about environmental monitoring, biodecontamination, equipment design and the evidence we use to support sterility assurance.
One question kept returning: what information do we need to understand whether our contamination controls are working, and what will we do with that information?
Having worked both as a QP and as an inspector, I approach that question from two closely connected perspectives. What would I need to know to support a batch certification decision? And what evidence would give me confidence that the process is consistently controlled?
What happens before the isolator doors close
Much of the opening debate focused on whether personnel and the Grade C background environment should be monitored during isolator loading and setup, before the bio decontamination cycle. Several participants questioned whether this monitoring was necessary.
My position was that, where Grade C monitoring is normally performed weekly and isolator loading also takes place once a week, the monitoring program should capture the loading activity. From an environmental monitoring risk assessment perspective, opening the isolator and introducing equipment and personnel into it would reasonably represent the highest risk event in that area. Excluding that activity would therefore require a clear scientific and risk-based justification.
I also argued that personnel interventions into the isolator environment that will subsequently become Grade A (affectionately known as the ‘to be Grade A’ environment) should be monitored as a check on the contamination burden introduced during loading. Opening the doors, installing equipment, and completing setup create opportunities to introduce contamination. We need to understand that challenge, including the condition of the Grade C background environment and the microorganisms associated with personnel and loading activities.
A validated bio decontamination cycle provides important assurance. However, understanding the burden presented to that cycle helps us assess loading practices, personnel controls, and whether the challenge remains within the conditions the process has been demonstrated to address. We should still ask whether the challenge used during validation represents contamination that could be introduced during actual operations. Microorganisms carried on particles or protected within deposits may present a different challenge from the biological indicators used to demonstrate cycle effectiveness.
For me, knowing that contamination was present can be useful even where a subsequent cycle is intended to address it. Understanding the organism, its source, and the circumstances of its introduction may identify weaknesses in material transfers, cleaning, loading practices, personnel controls or management of the background environment.
Other panelists asked equally important questions. What decision would a pre-cycle monitoring result support? How would it be investigated? Does existing background monitoring provide the information needed? Those questions should shape the program. Collecting more data has value when it improves process understanding or enables meaningful action.
Monitoring has its own risks
The discussion also highlighted a practical concern: in a highly automated, low-intervention isolator, environmental monitoring can itself become one of the most frequent interventions.
Every manipulation needs consideration. The information gained from sampling should be weighed alongside the risks introduced by performing it. However, this does not negate the regulatory requirements to continuously monitor the Grade A environment, both with viable and particulate monitors in positions that give the data needed to make the right decision.
That makes sample location, timing, technique, and data quality particularly important. During my inspection career, I saw facilities reporting no growth in Grade A monitoring while still experiencing sterility failures. Absence of growth on a monitor does not demonstrate the absence of contamination in the environment; to borrow a favorite phrase of my colleague Andy Hopkins, “absence of evidence is not evidence of absence”.
We need to understand what our monitoring can detect, where its limitations lie, and how it contributes to the wider contamination control strategy.
Cleanability starts with design
A point I raised was the importance of cleaning access.
An isolator may offer sophisticated controls, but operators still need to reach and clean its surfaces effectively. If routine cleaning requires someone to put their knees or body inside the enclosure, that should prompt serious questions about the design and the operating method.
Access doors, equipment layout and reach should be considered during procurement and mock-up reviews. Features described as optional can become essential when the team has to perform the cleaning.
The same scrutiny applies to robotics. Repeatable movements and reduced human intervention offer clear benefits, but cables, seals, and inaccessible surfaces can create additional challenges. Automation needs to work as part of a system that can be cleaned, decontaminated, and operated reliably.
First air and the potential to add contamination
Another lively part of the debate concerned first air and how we interpret its protection of exposed sterile product and critical surfaces. The point raised was that we need to consider an interruption in terms of its potential to add contamination. Treating every object in the airflow path as presenting the same risk can obscure the question we actually need to answer.
The distinction matters. When a component, equipment surface, or intervention sits between the air supply and a critical location, we need to understand whether it could introduce contamination into the air reaching that location. Its condition, how it has been cleaned, sterilized or bio decontaminated, and how it is handled all contribute to that assessment. The potential to add contamination is the central concern.
That assessment also needs to consider what the obstruction does to the airflow itself. Even where an object is not an obvious source of contamination, its position or movement may affect the protection of a critical surface. We therefore need to examine both the potential for contamination to be introduced and whether protective airflow is maintained during actual operations.
The panel challenged interpretations that reduce first air to a simple rule that nothing can ever interrupt the airflow. Such an approach can lead to impractical design demands without clearly explaining the contamination risk being addressed. Equally, describing an item as decontaminated should not end the assessment: its suitability and the effect of its presence need supporting evidence.
For me, this brings the discussion back to process understanding. Airflow visualization under representative operating conditions, including relevant interventions, should help us understand how the equipment and activities affect protection at critical locations. The findings need to be considered alongside cleaning, bio decontamination, barrier integrity, and operational controls within the contamination control strategy.
First air is an important protective mechanism. The useful question is whether an interruption has the potential to add contamination or compromise the protection needed at the critical location, and what evidence supports our conclusion.
Biological indicators and why I prefer the term “BI survivor”
During the debate, I challenged the term “rogue BI”, preferring the coined term by Mr. Friedman of USFDA of “BI survivor”. Calling an indicator “rogue” risks dismissing an unexpected result before its cause has been established. “BI survivor” directs attention towards investigation: did spores survive the bio decontamination cycle, or is there another explanation, such as contamination during handling or testing?
I also questioned what explanations involving spore clumping or multiple layers tell us about the cycle’s effectiveness against real contamination. Microorganisms introduced during loading may be carried on particles or protected within deposits. If these conditions are used to explain a BI survivor, we should consider whether they reveal a relevant process limitation. This reinforces the importance of understanding the contamination burden introduced during loading and investigating positive BIs without prejudging the outcome.
The panel also discussed chemical and enzymatic indicators. I expressed support for exploring these technologies and the view that they may offer more reliable information in some applications. Other participants highlighted the value of enzymatic indicators during cycle development, particularly their ability to provide rapid, graduated information rather than only a growth or no-growth endpoint. There was support for gathering comparative evidence alongside BIs to establish where these alternatives could be used confidently.
For more on the discussion of the fragility of vaporized Hydrogen Peroxide decontamination cycles, please see previous blogs by Andy Hopkins.
Blogs by Andy Hopkins on the Lachman Blog: https://www.lachmanconsultants.com/author/ahopkins/
MHRA Blog: https://mhrainspectorate.blog.gov.uk/
For me, the common thread was clear: choose indicators that help us understand the process and investigate unexpected results using terminology that encourages inquiry rather than prejudges the answer. As a final thought; the use of an isolator with a validated decontamination cycle will always go a long way to produce a less risk environment and patient benefit than a RABS system without one.
Making room for different views
There were differing opinions about monitoring, first air, flexible wall isolators and alternatives to biological indicators. Those differences made the discussion useful.
The panelists brought experience from different equipment designs, applications and operating environments, points raised that flexible wall isolators are outdated and problematic when it comes to integrity testing of the unit itself.
We also discussed the importance of questioning expectations while recognising that regulatory perspectives may be informed by contamination events and inspection experience that are not widely visible to the wider pharmaceutical industry.
The closing question was whether anything would change our minds.
An answer was given that good data should always make us willing to reconsider. That includes trend data, representative sampling, and confidence in how the information was generated. I would like to emphasise the ‘good’ data aspect of that answer, as not all data is equal, and poorly placed monitors will give a false sense of security.
I left the discussion with the same priority I brought into it: understanding how contamination can enter the process, how effectively it is controlled, and whether the evidence supports the assurance we claim. Design is key. No amount of monitoring data can support a poorly designed environment or process.
That is a debate worth continuing.